100 % original. Más de 100 000 piezas en stock. Listas para enviar.

  • es

The Bently Nevada 330130-045-00-05, also cataloged as the 330130 Extension Cable, operates as a dedicated hardware component for RF signal transmission within 3300 XL Proximitor Sensor networks. It acts as the electrical interface linking a proximity probe to a remote transducer module, transferring raw high-frequency displacement signals back to the monitoring system chassis without inducing uncompensated loop impedance shifts or phase errors.

Suffix Breakdown & Model Matrix

  • 330130 - Base Model: 3300 XL Standard Extension Cable
  • -045 - Cable Length Option: 4.5 metres (14.8 feet)
  • -00 - Connector Protector and Cable Option: Standard cable (non-armored)
  • -05 - Agency Approval Option: Multiple Approvals (ATEX, IECEx, CSA, etc. for hazardous locations)

Hardware Specifications

Parameter Specification
Model 330130-045-00-05
Brand Bently Nevada
Origin United States
Weight 0.23 kg nominal
Dimensions 4.5 m physical length (75 ohm coaxial geometry)
Operating Temp -51 to +177 deg C
Power Consumption 0.0 W (Passive component)
System Calibration 3300 XL 5 mm and 8 mm Transducer Systems
Nominal Capacitance 69.9 pF/m (21.3 pF/ft)
Minimum Bend Radius 25.4 mm (1.0 in)

Eddy-Current Signal Shielding & Cross-Talk Suppression

The 330130-045-00-05 coaxial structure provides cross-talk suppression and maintains strict calibration constants for eddy-current probe scaling. By securing continuous shielding over the 4.5 m path, the cable protects the low-voltage dynamic signals from electromagnetic noise generated by nearby machine drivers. This stability supports accurate gap voltage validation targeting nominal -10 VDC levels, protecting the linearity of the static and dynamic data passed to the continuous monitoring hardware.

Frequently Asked Questions

Q: Can this standard non-armored extension cable be cut or spliced in the field to match short routing runs?

A: No. Altering the physical length of the cable modifies its internal capacitance and inductance parameters. Splicing introduces impedance anomalies that degrade the linear scale factor of the 3300 XL system, invalidating machinery displacement measurements.

Q: How does the "Multiple Approvals" (-05 suffix) impact the hardware configuration compared to the base version?

A: The physical dimensions and electrical properties remain identical. The -05 designation verifies that the cable has undergone testing and marking to comply with specific intrinsic safety parameters when connected to certified barrier devices in classified hazardous zones.

Field Installation Guidelines

  • Maintain the standard minimum bend radius of 25.4 mm during cable tray layout execution to avoid dielectric deformation.
  • Ensure all terminal connections are torqued according to manufacturer specifications to avoid loose contacts that shift baseline loop resistance.
  • Ground the outer shield solely at the designated Proximitor sensor terminal block to prevent the creation of parasitic ground loops.
  • Isolate coaxial connectors from structural metal conduits by using the integrated insulation sleeves or external self-fusing tape.

¿Qué dicen nuestros clientes sobre nosotros?