Configured for shaft displacement and vibration measurement in rotating machinery monitoring networks, the Bently Nevada 330195-02-12-05-00 (330195 Reverse Mount Proximity Probe) provides direct physical/electrical execution. It operates by generating a localized eddy-current field at the probe tip to measure dynamic distances between the target shaft and the sensor, converting displacement into high-frequency voltage shifts processed by the downstream Proximitor sensor.
| Option Code | Specification Parameter | Value / Description |
|---|---|---|
| 330195 | Base Model | 3300 XL 8 mm Reverse Mount Probe (3/8-24 UNF threads) |
| -02 | Unthreaded Length Option | 0.2 in |
| -12 | Overall Case Length Option | 1.2 in |
| -05 | Total Length Option | 0.5 metre (1.6 feet) |
| -00 | Agency Approval Option | Not required |
| Parameter | Specification |
|---|---|
| Model | 330195-02-12-05-00 |
| Brand | Bently Nevada |
| Origin | USA |
| Weight | Approx 0.15 kg (Shipping Weight 2.0 kg) |
| Dimensions | 3/8-24 UNF reverse mount body with 0.5 m cable |
| Operating Temp | -51 deg C to +177 deg C (-60 deg F to +350 deg F) |
| Power Consumption | Passive transducer (driven by external Proximitor loop supply) |
| Thread Type | 3/8-24 UNF threads |
| Unthreaded Length | 0.2 in |
| Overall Case Length | 1.2 in |
| Total System Length | 0.5 metre (1.6 feet) |
| Pressure Sealing | Viton O-ring pressure seal (tip-to-case differential) |
| Agency Approval | Not required (Option 00) |
The 330195-02-12-05-00 proximity probe relies on precise eddy-current probe scaling to ensure linear voltage conversion over target operational distances. Installation requires strict gap voltage validation (-10 VDC targets) measured at the Proximitor output terminals to establish correct physical centering relative to the target shaft. Proper spatial mounting geometry must be enforced during reverse mount fitting to guarantee cross-talk suppression and eliminate mutual inductive interference between closely spaced probe tips.
Q: What is the primary function of the integrated Viton O-ring in this reverse mount configuration?
A: The Viton O-ring forms a fluid seal between the probe tip and casing, designed to hold differential pressure across machine housing walls in fluid-film bearing applications.
Q: How is the -10.0 VDC baseline verified during field setup?
A: Adjust the axial depth of the reverse mount probe using a threaded sleeve while measuring output voltage at the Proximitor terminals until a steady -10.0 VDC is observed at the midpoint of the linear range.
Our production line was down due to a faulty Bently Nevada monitor. We found it in stock here and they shipped it via DHL immediately. Back in operation within 48 hours—truly a lifesaver for our plant.
Finding brand new, original modules is critical for our safety. This team delivers factory-fresh components with full certification. Their commitment to only selling new products makes them our most trusted partner.
We have sourced Honeywell and Schneider parts here for two years. Every shipment arrives well-packaged with original factory seals intact. Clear communication and the best lead times in the industry.